Where does the UDI need to appear in device records?
By Savi C., Co-Founder · Published Nov 16, 2026
The label is only the start
Part 801 puts the UDI on the device. Under 21 CFR 801.20, the label of every medical device and every device package must bear a UDI that meets Part 801 and Part 830, with exceptions set out in the regulation. Under 801.40 it must appear in easily readable plain text and in automatic identification and data capture form, such as a barcode. Devices intended to be used more than once and reprocessed before each use must also carry a permanent direct marking under 801.45, unless an exception applies.
A UDI has two parts, defined in 21 CFR 830.3. The device identifier is the fixed portion that identifies the labeler and the specific version or model. The production identifier is the variable portion that carries the lot or batch, serial number, expiration date or manufacturing date, whichever of these appear on the label. This split matters for records: the device identifier is registered once in the GUDID, but the production identifier changes with every lot, so only your own records can tie a full UDI to a specific build.
Where the UDI has to appear
Record or submission
Record for each device or batch
Requirement
820.35(c): the UDI must be recorded for each medical device or batch, in addition to ISO 13485 clauses 7.5.1, 7.5.8 and 7.5.9
Record or submission
Labeling inspection
Requirement
820.45(a)(1): labeling and packaging examined for the correct UDI or UPC, or other device identification, before release or storage
Record or submission
Complaint records
Requirement
820.35(a)(3): any UDI or UPC and any other device identification, for complaints that are reportable, that must be investigated, or that were investigated anyway
Record or submission
Servicing records
Requirement
820.35(b)(2): any UDI or UPC and any other device identification of the device serviced
Record or submission
UDI system
Requirement
820.10(b)(1): a documented system to assign UDIs in accordance with Part 830, to meet ISO 13485 clause 7.5.8
Record or submission
Medical device reports
Requirement
803.52(c)(4): the UDI that appears on the device label or package, with model, catalog, serial or lot number and expiration date, if known
Record or submission
Corrections and removals reports
Requirement
806.10: the UDI on the label or package, or the device identifier, UPC or model and the lot or serial numbers
Record or submission
GUDID
Requirement
830.310: the device identifier and device attributes for each version or model, including the type of production identifiers on the label, updated under 830.330 when information changes
Record or submission
Labeler UDI records
Requirement
830.360: records of all UDIs used and the version or model for each device identifier, kept for 3 years after the version or model is no longer marketed
Where gaps usually show up
The production record and the complaint file are where the UDI most often goes missing. A traveler may record the lot and serial number but not the full UDI string, or a complaint intake form may capture a product name and a lot number typed from a photo. The QMSR made both explicit: 820.35(c) for the device or batch record and 820.35(a)(3) for complaints. A second common gap is drift between the GUDID entry, the approved label in the medical device file and the label actually printed, for example after a packaging quantity change that should have produced a new device identifier.
UDI record checks
Production and release
- The UDI recorded in the record for each device or batch, not only lot and serial number
- The device identifier on the printed label matches the one assigned to this version or model
- Production identifier values on the label match the lot, serial and dates in the record
- Labeling inspection records confirm the UDI was checked before release
Post-market records
- Complaint records for reportable or investigated complaints include any UDI or UPC
- Servicing records include any UDI or UPC of the device serviced
- MDR and correction or removal reports carry the UDI from the label or package, where known
Master data
- GUDID entries match the approved label content in the medical device file
- GUDID updated when information changes, within the timing 830.330 sets
- A retained list of every UDI used and its version or model, as 830.360 requires
How Litewave fits
Checking the UDI is a cross-record comparison: label, traveler, inspection record and, later, complaint file. Litewave reads paper or electronic records, checks every entry against your master record and specifications, and shows QA only the exceptions, each linked to the page it came from. Litewave proposes, qualified people approve. Our published results come from pharmaceutical manufacturing, and device figures will be published only when a device customer has agreed to them.
Questions we get asked
Does the GUDID hold lot and serial numbers?
No. Under 830.310 the labeler submits the device identifier and attributes of each version or model, including which types of production identifiers appear on the label, but not the production identifier values themselves. Those live only in your production and distribution records.
Does the full UDI have to be recorded, or is the lot number enough?
820.35(c) says the UDI must be recorded for each medical device or batch. Recording only the lot or serial number leaves the device identifier to be inferred, which is the gap the QMSR text was written to close.
Who is responsible for the UDI, the manufacturer or the labeler?
Part 830 places UDI and GUDID obligations on the labeler, the person who causes the label to be applied or modified for commercial distribution. That is usually the brand owner. Contract manufacturers still need the UDI in their own records when their quality agreement makes them responsible for labeling or production records.
Sources
- 21 CFR 801.20, Label to bear a unique device identifier (eCFR)
- 21 CFR 801.40, Form of a unique device identifier (eCFR)
- 21 CFR 801.45, Devices that must be directly marked (eCFR)
- 21 CFR 830.3, Definitions (eCFR)
- 21 CFR 830.20, Requirements for a unique device identifier (eCFR)
- 21 CFR 830.300, Devices subject to device identification data submission requirements (eCFR)
- 21 CFR 830.310, Information required for unique device identification (eCFR)
- 21 CFR 830.330, Times for submission of unique device identification information (eCFR)
- 21 CFR 830.360, Records to be maintained by the labeler (eCFR)
- 21 CFR 820.10, Requirements for a quality management system (eCFR)
- 21 CFR 820.35, Control of records (eCFR)
- 21 CFR 820.45, Device labeling and packaging controls (eCFR)
- 21 CFR 803.52, Individual adverse event report contents (eCFR)
- 21 CFR 806.10, Reports of corrections and removals (eCFR)
