What changed for DHRs, DMRs and DHFs under the QMSR
By Savi C., Co-Founder · Published Nov 16, 2026
Old terms and what replaced them
Old QSR term
Device master record (DMR)
Under the QMSR and ISO 13485
Medical device file, clause 4.2.3
Old QSR term
Device history record (DHR)
Under the QMSR and ISO 13485
Record for each device or batch, clause 7.5.1
Old QSR term
Design history file (DHF)
Under the QMSR and ISO 13485
Design and development file, clause 7.3.10
What stayed the same
The substance of each record survives. You still need a record for each device or batch that provides traceability and identifies the quantity manufactured and the quantity approved for distribution, and you still need to show the device was built to its approved specifications.
The QMSR also keeps some FDA-specific requirements on top of ISO 13485, including labeling and packaging controls under 820.45, which require labeling to be inspected before release for the correct UDI, expiration date, and storage and handling instructions.
What changed in practice
Management review, internal audit and supplier audit reports are no longer off limits. The old 21 CFR 820.180(c) exempted them from FDA inspection, and the QMSR did not carry that exemption forward. Those reports should now be written as if an investigator will read them.
Inspections changed too. FDA retired the QSIT inspection technique and now inspects under Compliance Program 7382.850, against ISO 13485 as incorporated by the QMSR. The same standard your notified body and MDSAP auditor use now applies to FDA inspections as well.
What to update in your quality system
Documents and terminology
- Map each procedure that references the DMR, DHR or DHF to its ISO 13485 equivalent
- Decide whether to keep the familiar terms internally, and say so in your quality manual
- Check that your per-device or per-batch record captures traceability and both quantities required by clause 7.5.1
Inspection readiness
- Bring management review, internal audit and supplier audit reports up to inspection standard
- Confirm labeling inspection before release is recorded as 820.45 requires
- Run a gap assessment against ISO 13485 as incorporated, not only against your certificate
Questions we get asked
Can we keep calling it a DHR?
Yes. The regulation no longer uses the term, but nothing stops you using it internally as long as the record meets ISO 13485 clause 7.5.1. Many manufacturers have kept the name and noted the mapping in their quality manual.
Can FDA now read our internal audit reports?
Yes. The QMSR removed the exemption in the old 21 CFR 820.180(c), so management review, internal audit and supplier audit reports can be inspected.
Is ISO 13485 certification enough for the QMSR?
No. Certification helps, but FDA inspects against the regulation, which keeps requirements beyond ISO 13485, and a certificate is not evidence that you meet them.
